Calibration out of tolerance: what to do next.

The cal cert comes back with "as found: out of tolerance" next to a micrometer that measured parts for six months. This is the procedure for working out which of those results still stand, with a worked example, a decision table, and how to set the calibration tolerance in the first place.

The short answer
  • When a gauge is found out of tolerance at calibration, quarantine it, read the as-found error, list every result it produced since its last good calibration, and recheck each result against its tolerance with that error applied. Disposition suspect product through an NCR, notify customers if suspect parts shipped, and keep the assessment as a record.

What does "out of tolerance" mean on a calibration report?

Out of tolerance (OOT) on a calibration report means the instrument's as-found reading at one or more test points was outside the acceptance limits set for that instrument. "As found" is the reading taken before the lab adjusts or repairs anything. "As left" is the reading after.

As-found data describes the gauge you were using on parts, so it drives the look-back. As-left data describes the gauge from now on. A cert that says "adjusted, now in tolerance" with no as-found values leaves you nothing to assess. Ask your lab for as-found data on every calibration.

An OOT result is not a bad part. A micrometer reading 0.0003" high matters for a ±0.0002" feature and not for a ±0.005" one. The procedure below tells them apart, in writing.

What is the difference between out of calibration and out of tolerance?

Out of calibration usually means the gauge is past its calibration due date. Out of tolerance means it was calibrated and failed. Shops use "out of cal" loosely for both; keep the terms separate in NCRs and audit answers.

Out of calibration (overdue)Out of tolerance (failed cal)
What happenedThe cal due date passed and the gauge stayed in use.The lab measured the gauge and at least one as-found point was outside its acceptance limits.
What you knowNothing yet. The gauge may be fine. There's no evidence either way.The size and direction of the error, at the test points the lab checked.
Results in questionEvery reading taken after the due date.Every reading taken since the last record showing the gauge was in tolerance.
First actionPull the gauge and send it for calibration. The as-found result decides what happens next.Quarantine the gauge and start the impact assessment.
If the as-found result passesThe readings were very likely valid. Record that and fix the recall process that let it lapse.Not applicable.
If the as-found result failsTreat it as an OOT and run the full procedure, back to the last good cal, not just the due date.Run the full procedure.
Audit viewUsing an overdue gauge on an acceptance record is a finding by itself, even if the gauge later passes.An OOT is not a finding. Having no documented assessment of earlier results is.

The overdue case is the one that bites in audits. An auditor picks one FAI and one line on Form 3: a bore checked with a pin gauge. The pin gauge's cal had lapsed six weeks before the inspection. Now every part that gauge measured since it went overdue is in question. If the pin gauge then fails its as-found check, the window stretches back further, to its last good calibration.

What does AS9100 require when a gauge is found out of tolerance?

AS9100D clause 7.1.5.2 (measurement traceability) requires you to determine whether the validity of previous measurement results has been adversely affected when measuring equipment is found unfit for its intended purpose, and to take appropriate action. The same clause asks for a recall process for equipment needing calibration, and a register of equipment that records each gauge's calibration method, frequency and acceptance criteria.

Clause 8.7.1 then applies to any product the assessment shows is nonconforming. For product already shipped, AS9100D asks for timely reporting of nonconformities affecting delivered products to the customer. Medical shops have the same expectation under ISO 13485:2016 clause 7.6: assess and record the validity of previous results, and act on the equipment and any product affected.

The standards don't say how to do the assessment, only that you have a process and a record of each one. Clause wording here is paraphrased; check your own copy.

What is the calibration out of tolerance procedure, step by step?

A calibration out of tolerance procedure has eight steps: quarantine, read the as-found data, set the period at risk, find every use, assess each result, disposition, notify, and close out with records.

  1. Quarantine the gauge. Tag it out of service, even if the lab has already adjusted it. Check sister gauges that might share the cause.
  2. Read the as-found data. Note the error at each test point, its direction (reads high or low), and whether it is a consistent offset or grows across the range. Note the lab's measurement uncertainty. If the cert only says "OOT", call the lab and get the numbers before going further.
  3. Set the period at risk. Start at the last record showing the gauge was in tolerance: usually the previous calibration's as-left result, or a later logged intermediate check that passed. End at the date the gauge was pulled from use. With no records in between, assume the error existed for the whole interval.
  4. Find every use in that period. Every job, lot, FAI and characteristic where that gauge ID appears on an inspection record. Include any other gauge or master the suspect gauge was used to set or verify, such as a bore gauge set from the micrometer. Those are suspect too.
  5. Assess each result against its tolerance. Apply the as-found error to the recorded reading and compare the corrected value to the drawing limits. Correct by the whole as-found error, not just the amount it exceeded the calibration tolerance. The worked example below shows this.
  6. Disposition what's affected. Open an NCR for every suspect lot or FAI characteristic. Re-inspect retained parts with a good gauge where you can. Material review decides use-as-is, rework or scrap, with design authority or customer approval where your contract requires it.
  7. Notify where required. If suspect parts shipped, tell the customer, using whatever notice their PO or supplier quality manual requires. Name the parts, lots, serial numbers and characteristics, and what you're doing about them.
  8. Close it out. File everything together (see records). Fix the cause of the drift, and shorten the interval if the gauge type keeps failing.

Step 4 is where most shops get stuck. If inspection records say "micrometer" instead of a gauge ID, you can't tell which of five micrometers measured a part. The honest result is then that every part measured with any micrometer in that period is suspect. Recording the gauge ID against every reading is what makes the look-back a one-hour job instead of a week.

What does an out of tolerance impact assessment look like? A worked example

An impact assessment applies the gauge's as-found error to each recorded reading and checks whether the corrected value is still inside the drawing tolerance. Here is one, start to finish, for a 0–1" outside micrometer.

The gauge. Micrometer M-014, 0–1", 0.0001" graduation. The register lists its calibration tolerance as ±0.0001", which matches what makers typically specify for this size.

The as-found data. Calibrated 2026-09-15. As found, it read +0.0003" at every test point from 0 to 1.000", a consistent offset of the kind a knock to the zero setting produces. The lab reset the zero, and the as-left result is within ±0.0001". The previous calibration, on 2026-03-10, was in tolerance as left. There are no logged checks in between.

The period at risk. 2026-03-10 to 2026-09-15. A search of inspection records for M-014 in that window finds five characteristics across five jobs.

The logic. The micrometer read 0.0003" high, so the true size of every part was 0.0003" smaller than recorded. Subtract 0.0003" from each reading. Readings near the upper limit get safer. Readings near the lower limit are the ones at risk.

Job / recordCharacteristicLimits (in)Recorded range (in)Corrected range (in)Result
FAI 4471-102 Rev D, balloon 12Shaft OD .3750 ±.0005.3745 – .3755.3750 – .3753.3747 – .3750No impact. Lowest corrected value is .0002 inside the limit.
Job 1182, part 2209-04, 40 pcsOD .6250 +.0000/-.0010.6240 – .6250.6242 – .6249.6239 – .6246Affected. Parts recorded at .6242 are actually .6239, .0001 undersize. NCR, re-inspect, notify if shipped.
FAI 3310-200 Rev A, balloon 4Thickness .250 ±.005.2450 – .2550.2480 – .2510.2477 – .2507No impact. The error is 3% of the tolerance band.
Job 1207, part 5120-11, 200 pcsSpacer thickness .0625 ±.0002.0623 – .0627.0624 – .0627.0621 – .0624Affected. Up to .0002 undersize. NCR. Also: a ±.0002 band is too tight for this gauge (see below).
Job 1215, part 6602-03, 25 pcsPin OD .4375 +.0005/-.0000.4375 – .4380.4379 – .4380.4376 – .4377No impact. Readings were near the top limit, so the correction moves them toward the middle.

Three of five results stand, each with a one-line reason. Two need NCRs. Unless each of the 40 parts in job 1182 was recorded, treat the lot as suspect and re-inspect what you still hold. If FAI 3310-200 had been affected, the FAI itself would need an NCR reference against balloon 4 and a re-measurement with a good gauge. Check with the customer before deciding whether the corrected FAI must be resubmitted.

Two refinements matter in real assessments:

  • Leave margin for uncertainty. A corrected value on the limit isn't a clear pass. Set one rule, such as "suspect if within the cal uncertainty of a limit", and apply it to every line.
  • Use the error at the right point. If the error grows across the range instead of staying constant, use the as-found error at the test point nearest each measured size, or the worst case if you can't tell.

Every line in that table starts with a balloon number. Balloon the drawing first, so each result has one.

Try the free ballooner

How do you find every job a gauge was used on?

You find every job a gauge measured by searching inspection records for its unique gauge ID within the period at risk. That only works if the ID is on the record. Where to look:

  • FAI reports. The gauge recorded against each Form 3 result. See the Form 3 guide for where results and references go.
  • In-process and final inspection sheets. First-piece checks, travelers, and lot acceptance records.
  • Other gauges it set. A bore gauge or height master set from the suspect gauge, and every job those touched.

If you can't tell which jobs, the scope is every job in the period that used that type of gauge. A broad, documented scope beats a narrow, guessed one. The AS9100 audit checklist for FAI records covers how auditors trace a gauge from a sampled FAI.

How do you determine a calibration tolerance for a gauge?

Determine a calibration tolerance by starting from the manufacturer's accuracy specification or a recognized standard for that gauge type, then checking it is tight enough for the part tolerances the gauge will measure. Write the result in your equipment register as the acceptance criteria. AS9100D asks the register to hold them.

  1. Manufacturer's specification. A typical 0–1" mechanical outside micrometer is specified to ±0.0001" by its maker. Mitutoyo's 103 series is one example. Start there.
  2. Published standards. In the US, ASME B89.1.13 covers micrometers, including maximum permissible errors for outside micrometers; Mitutoyo America bases its calibration method on it. Internationally, ISO 3611 covers external micrometers, and its 2023 edition adds default maximum permissible errors. Take the numbers from your own copy.
  3. Fit for the work. Compare the gauge's accuracy with the tightest tolerance it measures. A common shop rule of thumb is 10:1: gauge accuracy, or resolution, about one tenth of the part's total tolerance band. A 4:1 ratio is often treated as the practical minimum. This is a rule of thumb, not a standard.
  4. Tighten for critical use. If one micrometer is reserved for tight work, you can give it a tighter calibration tolerance than the maker's spec, provided the lab can verify it.

In the worked example, the spacer toleranced ±.0002" has a total band of .0004". A micrometer accurate to ±.0001" uses half of that band on its own error. By the 10:1 rule the spacer needed a gauge good to about .00004", which means a bench comparator or an indicator against gauge blocks, not a hand micrometer. The OOT just exposed a gauge selection problem that was there all along.

Don't confuse your gauge's calibration tolerance with the lab's TUR. The test uncertainty ratio (TUR) compares your gauge's tolerance with the lab's measurement uncertainty, and it tells you whether the lab's pass/fail call can be trusted. ANSI/NCSL Z540.3 limits the probability of a false accept to 2%, and a TUR of at least 4:1 is the widely used fallback where that probability isn't calculated. If your lab reports a TUR below 4:1 on a gauge you use for tight work, ask how they made the call.

What records should an OOT investigation leave behind?

An OOT investigation should leave one file per event that an auditor can read without asking you to explain it. A good file holds:

  • The calibration certificate with as-found and as-left data, and the lab's uncertainty.
  • The period at risk and how you set its start date.
  • The list of every job, lot, FAI and characteristic searched, including the ones with no impact.
  • The corrected-value assessment for each, like the table above, and the decision rule you used.
  • The NCR numbers, dispositions and any customer approvals.
  • Customer notifications, with dates.
  • The cause, the corrective action, and any change to the calibration interval.

The "no impact" lines matter as much as the affected ones: they're the evidence clause 7.1.5.2 asks for. If a CMM fails verification, the same steps apply to every program run in the window; see how to read a CMM report. For the FAI side, see why FAI reports get rejected.

FAQ

What does out of tolerance mean on a calibration report?

Out of tolerance (OOT) on a calibration report means that at one or more test points the instrument's as-found reading, taken before any adjustment, was outside the acceptance limits set for that instrument. The as-found values tell you how wrong the gauge was while you were using it. The as-left values tell you its condition after adjustment, for future use.

What is the difference between out of calibration and out of tolerance?

Out of calibration usually means the gauge is past its calibration due date, so nobody has verified it recently; it may still be accurate. Out of tolerance means a calibration was done and the gauge failed its acceptance limits. An overdue gauge is a gap in evidence. An out-of-tolerance gauge is a known error, and every result it produced since its last good calibration has to be assessed.

How far back do you go after a gauge is found out of tolerance?

Go back to the last point where you have a record showing the gauge was in tolerance. That is usually the previous calibration's as-left result. If you log intermediate checks, such as a daily check against a gauge block, the look-back starts at the last passing check. Without records in between, assume the error existed for the whole interval.

Do I have to tell my customer when a gauge fails calibration?

Not for the failed calibration itself. You must notify the customer when the impact assessment shows that nonconforming or suspect product may have been delivered. AS9100D clause 8.7.1 calls for timely reporting of nonconformities affecting delivered products to the customer, and many purchase orders set their own notice period. If the assessment shows no affected product, record that conclusion and keep it.

How do you determine a calibration tolerance for a gauge?

Start from the manufacturer's accuracy specification or a recognized standard such as ASME B89.1.13 or ISO 3611 for micrometers. Then check the result against the tightest part tolerance the gauge measures; a common rule of thumb is that gauge accuracy should be about one tenth of that tolerance. Tighten the limit for critical work, and write it in the equipment register.

Where does software help with a calibration recall?

Software helps most with step 4, finding every use, and with stopping the overdue case. Toolrick tracks gauges with their calibration dates, and a reading taken with an out-of-calibration gauge blocks sign-off rather than just showing a warning. Every reading is tied to the gauge that took it, so when a gauge fails you can find the FAIs it measured. The as-found assessment, the dispositions and the customer call are still yours. Start with the free ballooner, no account needed.

Related: AS9100 audit checklist: FAI records · AS9102 Form 3 explained · How to read a CMM report · Free AS9102 templates

Sources

  1. Why is OOT important in measurement traceability? (AS9100D 7.1.5.2, validity of previous measurements), Utah MEP
  2. AS9100 7.1.5 Monitoring and measuring resources (recall process, equipment register, acceptance criteria), a published QMS procedure excerpt
  3. AS9100 8.7 Control of nonconforming outputs (timely reporting of delivered nonconformities), a published QMS procedure excerpt
  4. ISO 13485 calibration requirements: assess and record validity of previous measuring results, Elsmar Cove forum
  5. Calibration out of tolerance, part 2 (as-found review, when the instrument drifted, impact on product), Beamex
  6. Why a 4:1 TUR is not enough (ANSI/NCSL Z540.3, 2% probability of false accept), Morehouse Instrument Company
  7. Outside micrometer calibration: how to calibrate (method based on ASME B89.1.13), Mitutoyo America
  8. BS EN ISO 3611:2023, design and metrological characteristics of micrometers for external measurements, BSI Knowledge

Know which FAIs a gauge touched.

Balloon the drawing free, then record readings against tracked gauges in a 14-day trial.