Gage R&R study: a worked example, by hand.

Most Gage R&R explanations stop at the formulas. Here is a full average-and-range study on a turned diameter: 90 real-looking readings, every intermediate number, and what the result means for a small shop deciding whether a micrometer is good enough for the job.

The short answer
  • A Gage R&R study measures how much of your measurement spread comes from the gauge (repeatability, EV) and from the people using it (reproducibility, AV). Typically 10 parts × 3 operators × 2–3 trials. AIAG guidance: %GRR under 10% is acceptable, 10–30% maybe, over 30% not, and ndc should be at least 5.

What does Gage R&R mean?

Gage R&R stands for gauge repeatability and reproducibility. A Gage R&R study splits the variation in a set of measurements into the part-to-part variation you want to see and the measurement variation you don't.

  • Repeatability (EV, equipment variation): the spread when one operator measures the same part several times with the same gauge. It is mostly the gauge, the fixture and the measuring technique.
  • Reproducibility (AV, appraiser variation): the difference between operators' averages when they measure the same parts. It is mostly technique, training and how the gauge is read.
  • GRR: the two combined, as √(EV² + AV²).

"Gage" and "gauge" mean the same thing; the AIAG Measurement Systems Analysis reference manual (MSA, 4th edition, 2010) spells it "gage". The manual is guidance, not a standard, and describes three methods for a variable study: range, average-and-range, and ANOVA. This post works average-and-range by hand, because every number can be checked with a calculator.

How do you set up a Gage R&R study?

Pick parts that cover the real process spread, pick the operators who actually use the gauge, and have each operator measure each part 2 or 3 times in random order without seeing earlier readings. The usual layout is 10 parts × 3 operators × 3 trials, which gives 90 readings.

  1. Parts. Take 10 parts from production that span the process range, not 10 consecutive parts that are nearly identical. Number them where the operators can't see the number.
  2. Operators. Use the 2 or 3 people who measure this feature on a normal shift. A study run only by the quality manager tells you little about second shift.
  3. Randomize. Present the parts in a different random order for each trial, so nobody remembers the last reading.
  4. Record to the gauge's full resolution. Don't round 12.0035 to 12.004 in your head. Rounding hides repeatability.

What does Gage R&R data look like?

Here is the example used through the rest of the post. The feature is a turned diameter of 12.000 ±0.025 mm (tolerance 0.050 mm), measured with a 0–25 mm digital micrometer that reads to 0.001 mm. Operators A, B and C each measured 10 parts three times.

PartA1A2A3B1B2B3C1C2C3Part avg
112.01712.01712.01612.01912.01612.01712.01912.01612.01512.0169
211.99611.99411.99111.99911.99711.99611.99411.99311.99311.9948
312.00512.00212.00212.00412.00212.00312.00212.00311.99912.0024
412.00112.00112.00112.00311.99912.00212.00012.00012.00012.0008
512.00512.00712.00412.00512.00712.00612.00712.00412.00612.0057
611.99111.98811.98911.99311.99211.99011.98811.98911.98811.9898
711.99711.99711.99711.99511.99711.99811.99611.99511.99711.9966
811.99511.99011.99411.99411.99611.99711.99611.99511.99511.9947
911.99211.99111.99411.99311.99211.99411.99211.99111.99211.9923
1011.99211.99311.99111.99411.99611.99411.99611.99111.99411.9934

For each operator, take the range of the three trials on each part (largest minus smallest), then average those 10 ranges. Also average all 30 of that operator's readings.

OperatorRanges, parts 1–10 (µm)Average range R̄Operator average X̄
A1, 5, 3, 0, 3, 3, 0, 5, 3, 22.5 µm11.99833
B3, 3, 2, 4, 2, 3, 3, 3, 2, 22.7 µm11.99967
C4, 1, 4, 0, 3, 1, 2, 1, 1, 52.2 µm11.99820

From here on, working in micrometres (µm, 0.001 mm) keeps the zeros out of the way.

What is the Gage R&R formula?

In the AIAG 4th edition average-and-range method, each component is a standard deviation: a range multiplied by a constant (K1, K2 or K3) that depends on how many trials, operators or parts were used. The 4th edition dropped the 5.15 multiplier the 3rd edition built into its constants, so don't mix constants from the two editions.

QuantityFormulaConstant used here
EV (repeatability)R̄̄ × K1, where R̄̄ is the average of the operators' R̄K1 = 0.5908 (3 trials; 0.8862 for 2)
AV (reproducibility)√[(X̄diff × K2)² − EV² ÷ (n × r)], where X̄diff is the largest operator average minus the smallest, n is parts and r is trialsK2 = 0.5231 (3 operators; 0.7071 for 2)
GRR√(EV² + AV²)
PV (part variation)Rp × K3, where Rp is the range of the part averagesK3 = 0.3146 (10 parts)
TV (total variation)√(GRR² + PV²)
%GRR100 × GRR ÷ TV (likewise %EV, %AV, %PV)
%GRR of tolerance100 × GRR ÷ (tolerance ÷ 6)
ndc1.41 × PV ÷ GRR, truncated to a whole number

K3 for other part counts: 2 parts 0.7071, 3 parts 0.5231, 4 parts 0.4467, 5 parts 0.4030, 6 parts 0.3742, 7 parts 0.3534, 8 parts 0.3375, 9 parts 0.3249. The EV² term inside AV takes out the share of the operator-average difference that is really just repeatability. If the number under that square root comes out negative, AV is set to 0.

How do you calculate Gage R&R? The worked example

Plug the summary numbers into the formulas in order. Each step uses only the results above it.

  1. Average range: R̄̄ = (2.5 + 2.7 + 2.2) ÷ 3 = 2.467 µm.
  2. EV: 2.467 × 0.5908 = 1.457 µm.
  3. Operator difference: X̄diff = 11.99967 − 11.99820 = 0.001467 mm = 1.467 µm (operator B highest, C lowest).
  4. AV: (1.467 × 0.5231)² = 0.7672² = 0.5886. EV² ÷ (10 × 3) = 1.457² ÷ 30 = 0.0708. √(0.5886 − 0.0708) = √0.5178 = 0.720 µm.
  5. GRR: √(1.457² + 0.720²) = 1.625 µm.
  6. Part range: Rp = 12.0169 − 11.9898 = 0.0271 mm = 27.11 µm (part 1 largest, part 6 smallest).
  7. PV: 27.11 × 0.3146 = 8.529 µm.
  8. TV: √(1.625² + 8.529²) = 8.683 µm.
ResultValue% of TV% of tolerance (0.050 mm)
EV, repeatability1.457 µm16.8%17.5%
AV, reproducibility0.720 µm8.3%8.6%
GRR1.625 µm18.7%19.5%
PV, part variation8.529 µm98.2%
ndc1.41 × 8.529 ÷ 1.625 = 7.40, truncated to 7

The % of tolerance column is 100 × 6 × component ÷ 50 µm. For GRR: 6 × 1.625 = 9.75 µm, and 9.75 ÷ 50 = 19.5%.

The percentages don't add up, and they shouldn't. %EV (16.8) plus %AV (8.3) is 25.1, not 18.7, because standard deviations combine as a root sum of squares. Likewise %GRR and %PV add to well over 100%. The squares add: 18.7² + 98.2² ≈ 100².

Two sanity checks are worth doing before trusting the result. First, the range chart: with 3 trials, the AIAG range-chart upper limit is D4 × R̄̄ = 2.574 × 2.467 = 6.35 µm. The largest single range in the study is 5 µm, so no operator had a wild reading. Second, ANOVA: the same 90 readings run through the ANOVA method give a %GRR of about 21%, a % of tolerance of about 20% and an ndc of 6. ANOVA also separates out the operator-by-part interaction, which the average-and-range method can't see, so expect small differences between the two methods.

What are the Gage R&R acceptance criteria?

The AIAG MSA manual gives %GRR bands as guidelines: under 10% is generally acceptable, 10% to 30% may be acceptable depending on the application, and over 30% is generally unacceptable. ndc should be 5 or more. These are guidance, and a customer's own requirement overrides them.

%GRRAIAG guidanceWhat a shop usually does
Under 10%Generally acceptableUse the gauge. File the study.
10% to 30%May be acceptable, based on how important the measurement is, the cost of a better gauge and the risk of a wrong callUse it with a documented decision, often with customer agreement. Watch parts measured near the limits.
Over 30%Generally unacceptable; improve the measurement systemDon't use it for this tolerance. Find the cause (next section).
ndc ≥ 5The measurement system can tell enough part groups apartAn ndc under 5 with a good %GRR usually means the study parts were too alike.

Our example lands at 18.7% of total variation and 19.5% of tolerance, with ndc 7: in the "may be acceptable" band on both measures, with enough discrimination. For a ±0.025 mm diameter that isn't a critical or key characteristic, most quality engineers would accept the micrometer, write down why, and keep an eye on readings within a few microns of a limit.

Which percentage to quote depends on the gauge's job. For pass/fail sorting, % of tolerance is the honest measure. For SPC or process adjustment, % of total variation matters, because a gauge can be fine against a wide tolerance and still blur a capable process. Many customers ask for both. Old reports may use 5.15 instead of 6 as the multiplier.

What should you do when a Gage R&R fails?

Look at which component is bigger. EV much larger than AV points at the gauge, fixture or part. AV much larger than EV points at the people and the method.

SymptomLikely causeWhat to try
EV large, AV smallGauge resolution too coarse, worn anvils, a loose fixture, part form error (out-of-round, taper) being picked up differently each timeA finer or stiffer gauge, a fixture or V-block, measuring at a marked location, checking form separately
AV large, EV smallOperators measure in different places, use different force, or read an analogue scale differentlyA written method with the measuring location, a ratchet or friction thimble for constant force, training, a digital readout
%GRR fine of tolerance, poor of TV, ndc lowThe study parts barely varyRe-run with parts that span the process, or report % of tolerance if that's how the gauge is used
Many zero ranges, readings in big stepsNot enough resolutionA gauge with at least ten times finer resolution than the tolerance (next section)

Moving the feature to a better instrument is also an option, though a CMM has its own sources of variation. See how to read a CMM report for the setup items that move every reading at once.

When does a small shop actually need a Gage R&R?

A shop needs a formal Gage R&R when a customer, a PPAP or an automotive standard asks for one. Otherwise it is good practice for tight tolerances and new gauges, not a blanket requirement.

  • Customer requirement. Purchase order clauses and supplier quality manuals often call for MSA on key characteristics. This is the most common trigger for aerospace and medical job shops.
  • AS9100D 7.1.5. The clause asks for measuring resources that are suitable and give valid and reliable results, with calibration, traceability and a recall process. It doesn't name Gage R&R. A study is one strong way to show a gauge is suitable for a tight tolerance when an auditor asks. Our AS9100 audit checklist for FAI records covers what auditors trace under 7.1.5.
  • AS9145 and PPAP. Aerospace APQP/PPAP (AS9145) and automotive PPAP include measurement system analysis studies in the submission. See PPAP vs FAI for how the two packages differ.
  • IATF 16949 7.1.5.1.1. Automotive suppliers must run statistical studies on each type of inspection, measurement and test system in the control plan, using methods from reference manuals such as the AIAG MSA, unless the customer approves others.
  • ISO 13485. Medical device customers often ask for test method validation, and a Gage R&R is a common part of it. Check the customer's procedure for the method and limits.

Without one of those triggers, spend study time where a wrong call is expensive: critical and key characteristics, and features where operators disagree.

What about attribute gages like go/no-go pins?

A go/no-go gauge gives pass or fail, not a number, so EV and AV can't be calculated. Attribute gauges are studied with an attribute agreement analysis instead. Operators repeatedly judge parts whose true status is known, including parts close to each limit, and the analysis reports how often they agree with themselves, each other and the reference (often summarized with kappa). The AIAG manual's attribute example uses 50 parts, 3 appraisers and 3 trials. A study made only of obviously good and obviously bad parts will always pass and proves nothing.

How does Gage R&R relate to first article inspection?

AS9102 Form 3 records a result for every characteristic, and that result is only as trustworthy as the gauge that produced it. A first article doesn't call for a Gage R&R study, but the gauge on each Form 3 row should be calibrated and capable of the tolerance it is checking.

  • The 10:1 rule of thumb. Pick a gauge whose resolution is no more than one tenth of the tolerance. For our 0.050 mm tolerance that means 0.005 mm or finer, so the 0.001 mm micrometer clears it. A caliper reading to 0.01 mm does not. The rule is a starting point; resolution is not the same as repeatability, which is what the study measures.
  • Calibrated is not the same as capable. Calibration checks the gauge against a reference standard; a Gage R&R checks the spread with your operators and parts.
  • A reading near a limit deserves a second look. With a GRR of 1.6 µm, a single reading of 12.024 on a 12.025 upper limit is not clearly in tolerance. Re-measure, or measure on a better instrument, before calling it.
  • Record the gauge ID. A reviewer who sees a result needs to trace it to a gauge, its calibration and, where one exists, its study. Missing gauge records are a common reason FAI reports get rejected. The Form 3 guide covers where that goes.

For true position, bolt circle and thread wire math, see the free shop calculators.

How do you do a Gage R&R in Excel?

Lay the readings out like the data table above, one row per part and one column per operator-trial, and the average-and-range method needs no paid template.

  1. Ranges: =MAX(B2:D2)-MIN(B2:D2) per part for each operator, then =AVERAGE() for R̄.
  2. Averages: =AVERAGE() per operator and per part, then =MAX()-MIN() for X̄diff and Rp.
  3. Constants in their own cells (K1, K2, K3, n, r, tolerance), so a 2-trial study only needs new constants.
  4. AV with a guard: =SQRT(MAX(0,(Xdiff*K2)^2-EV^2/(n*r))) returns 0 instead of an error.

Check a home-made sheet against the numbers above before using it for a customer. For ANOVA, use statistics software.

FAQ

What is an acceptable Gage R&R percentage?

The AIAG MSA manual gives guidance, not hard limits: %GRR under 10% is generally acceptable, 10% to 30% may be acceptable depending on the application, the cost of the gauge and the risk, and over 30% is generally unacceptable. The number of distinct categories (ndc) should be 5 or more. Your customer's requirements override these figures.

Should %GRR be calculated against total variation or tolerance?

Against total variation when the gauge is used to control the process, and against tolerance when the gauge is only used to sort good parts from bad. Many customers ask for both. %GRR of tolerance is 100 × 6 × GRR divided by (USL − LSL) under the AIAG 4th edition. Older reports used 5.15 instead of 6, so check which multiplier a report used before comparing.

What does ndc mean in a Gage R&R?

ndc is the number of distinct categories: how many separate groups of parts the measurement system can tell apart across the part spread in the study. It is 1.41 × PV ÷ GRR, truncated to a whole number. An ndc of 5 or more is the usual AIAG guidance. An ndc of 1 or 2 means the gauge can do little more than sort parts into high and low.

Does AS9100 require a Gage R&R study?

AS9100D does not name Gage R&R. Clause 7.1.5 requires measuring resources that are suitable and give valid, reliable results, plus calibration and traceability. A Gage R&R is one way to show suitability. Studies become a firm requirement when a customer flows them down, under an AS9145 PPAP, or under IATF 16949 7.1.5.1.1 in automotive work.

Is a calibrated gauge good enough without a Gage R&R?

Not always. Calibration checks the gauge against a reference standard in a lab. A Gage R&R checks how much the readings spread when your operators measure your parts on your floor. A calibrated micrometer can still be too coarse, or too operator-sensitive, for a tight tolerance.

Where software helps

Toolrick doesn't run Gage R&R studies; use a spreadsheet or statistics software for that. What it does is keep the gauge side of a first article honest. Each Form 3 reading carries the gauge that took it, and a reading taken with a gauge past its cal due date blocks sign-off until someone fixes it. Start by ballooning a drawing with the free ballooner, no account needed.

Related: How to read a CMM report · AS9100 audit checklist for FAI records · Why FAI reports get rejected · Shop calculators

Sources

  1. AIAG MSA-4:2010, Measurement Systems Analysis Reference Manual, 4th edition, AIAG (ANSI Webstore listing)
  2. 2.4. Gauge R & R studies, NIST/SEMATECH e-Handbook of Statistical Methods
  3. Variable Measurement Systems, Part 4: Gage R&R (average and range method, K1/K2/K3 tables), SPC for Excel
  4. Using the number of distinct categories in a gage R&R study, Minitab Support
  5. MSA statistics (%GRR guidance, ndc truncation), SigmaZone Quantum XL Help
  6. IATF 16949 clause 7.1.5.1.1: Measurement System Analysis, Advisera 16949Academy
  7. AS9100 calibration requirements (clause 7.1.5), Micro Precision Calibration

No out-of-cal readings on the FAIR.

Balloon the drawing free, then record each reading with its gauge in a 14-day trial.